Autocon Lab Instruments
01 / 02
Circulators (Baths)ACL-101

Circulating Cooling Bath

Dedicated Cooling Circulation for Laboratory Applications

The Autocon Circulating Cooling Bath delivers reliable sub-ambient cooling for external instruments and internal bath applications. Compact design with powerful refrigeration and a high-flow circulation pump.

ISO 9001:2015 CE Certified

Key engineering features

  • Sub-Ambient CoolingAchieves temperatures down to -20°C for demanding applications
  • High-Flow PumpPowerful circulation pump for consistent external cooling
  • Digital PID ControlPrecise set-point control with real-time temperature display
  • Compact FootprintSpace-efficient design ideal for crowded lab benches
  • Low MaintenanceSealed compressor system requires minimal maintenance
  • Drain ValveConvenient bottom drain for easy bath fluid replacement
Call sales
Models & capacities

Choose your size.

Standard factory models of the Circulating Cooling Bath. Anything outside these — chamber size, supply, controller — can be built to order.

ParameterModelACL-101-10ModelACL-101-25ModelACL-101-50
ConfigurationCirculating Cooling Bath (10L Compact)Circulating Cooling Bath (25L Standard)Circulating Cooling Bath (50L High Capacity)
Chamber capacity10 Liters25 Liters50 Liters
Peak power1.5 kW2.5 kW4.0 kW
SupplySingle Phase (230V)Single Phase (230V)Three Phase (415V)
Temperature / pressure-20°C to +30°C-30°C to +30°C-40°C to +30°C
Material of constructionSS 304 (Wetted parts)SS 316 / 304SS 316L (High Corrosion Resistance)
Dimensions320 x 420 x 580 mm420 x 520 x 720 mm580 x 680 x 920 mm
Add to quote
Specifications

Built to a standard.

Standard factory specification

ACL-101
01Temperature Range
-20°C to +40°C
02Temperature Stability
± 0.2°C
03Cooling Capacity
250W at 20°C / 150W at 0°C
04Pump Flow Rate
10–18 L/min
05Bath Material
Stainless Steel SS 304
06Control System
Microprocessor PID
07Power Supply
230V AC, 50 Hz

Specifications can be tailored to your facility. IQ/OQ/PQ protocols are available.

Electrical load

What will the Circulating Cooling Bath need from your supply?

50L
10 L compact300 L industrial
-20°C
−80 °C250 °C
8h
1 h24 h

Estimated peak power

2.5kW

Supply
Single phase, 230 V AC
Operating current
11 A
Recommended MCB
14A 1-pole
Daily consumption
13 kWh

Indicative values for planning. Custom transformers and power conditioning are available on request.

Utility & energy ROI

Calculate the 3-year running cost of the Circulating Cooling Bath.

Power savings, closed-loop water savings and payback period.

Open calculator
Configured together

Process Temperature Control & Reaction Suite

Engineered for rotary evaporators, reactors, and thermal cycling

External loop temperature control integrates with vacuum ovens for solvent removal, cold storage for reactant preservation, and stability chambers for shelf-life testing.

  • Ultra Low Deep FreezerACL-103

    Ultra Low Deep Freezer

    Freezers & Cold Storage

  • Vacuum OvenACL-111

    Vacuum Oven

    Ovens

  • Stability ChamberACL-108

    Stability Chamber

    Freezers & Cold Storage

Applications

  • Condensation Cooling,
  • Rotary Evaporators,
  • Reaction Cooling,
  • Sample Cooling,
  • Cold Traps